High end gas flow calibration made simple
Fluke Calibration 5141, 5142 and 5144 molbox RFM gas flow calibrator kits feature molbox RFM and molbloc-L configurations optimized to cover a very wide workload combined with molstic-L and other interconnect hardware needed for a complete calibration system. Simply supply 90 psi of pure N2 or air to the system. An included regulator delivers a stable regulated flow to the molbloc. Downstream of the molbloc, an included metering valve allows you to regulate the mass flow value required for the DUT. It’s that simple; supply 90 psi gas upstream, tare the reading, and adjust the fine metering valve to deliver the desired mass flow rate with an accuracy of ± 0.5 % of reading. Fluke Calibration gas flow calibrator kits eliminate complication, but don’t compromise molbox/molbloc best in class performance.
- Flow ranges cover from 1 sccm to 50 slm
- ± 0.5 % of reading total one-year accuracy
- Measures mass and volume flow with user settable reference pressure and temperature conditions
- Reference standard is upstream of the device under test, eliminating cross contamination and expensive damage to the flow elements
- Integrated gas flow regulation and adjustment hardware
- Expandable to over 5000 slm with additional molbloc flow elements and control hardware
- Includes traceable calibration for operation in N2 and Air, with corrections for other gases
- Gas supply and test adapters included for 1/4 in tube, 1/4 in NPT and 1/4 in BSP
- Includes advanced functions such as totalize, average, hi/lo, deviation, on-board purge, leak test and tare, available from the front control panel or by RS-232 and IEEE-488 remote interfaces
Measurement specifications |
|
Gas calibration included | Nitrogen (N2), Air |
Maximum working measurement line pressure | 600 kPa (87 psi) absolute |
Measurement range | <1sccm to 50 slm, depending on model. Flow ranges depend on test gas. Ranged indicated for N2 and Air. |
General specifications |
|
Power requirements | 85 V ac to 264 V ac, 47 Hz to 440 Hz, 18 VA max consumption |
Normal operating temperature range | 15 °C to 30 °C (59 °F to 86 °F) |
Storage temperature range | -20 °C to 70 °C (-4 °F to 158 °F) |
Vibration | Meets MIL-T-28800D |
molbox RFM weight and dimwnsions |
2.55 kg (5.6 lb) max 8 cm x 22.5 cm x 20 cm (3.1 in x 8.9 in x 7.9 in) approx |
molstic with molbloc weight and dimensions (max) |
9 kg (20 lb) 19 cm x 81.28 cm x 15.25 cm (7.5 in x 32 in x 6 in) approx |
Supply gas required |
99.998% pure N2 or Air regulated to 90 psig (600 kPa) Note: a fine pressure regulator is included, but an additional regulator (not included) may be required to reduce pressure supplied to the 514X kit to approximately 90 psi |
Gas calculations reported | Nitrogen (N2), Air, Argon (Ar), Carbon Monoxide (CO), Helium (He), Oxygen (O2), Carbon Dioxide (CO2), Carbon Tetrafluoride (CF4), Ethane (C2<;/sub>H6), Ethylene (C2H4), Fluoroform (CHF3), Hexafluoroethane (C2F6), Hydrogen (H2), Methane (CH4), Nitrous Oxide (N2O), Propane (C3H8), Sulfur Hexafluoride (SF6) |
Flow connections for gas supply and device under test | 1/4 in tube with adapters included for 1/4 in NOT and 1/4 in BSP |