商品特性:

Features

Fast, portable and easy to use

Correct target size for most thermometers

Calibration solutions from –15 °C to 500 °C (5 °F to 932 °F)

Radiometrically calibrated for traceable and consistent results

 

商品規格:

Specifications

Feature

4180

4181

Temperature range 
(@ 23 °C ambient, 
0.95 emissivity)

–15 °C to 120 °C

35 °C to 500 °C

Total accuracy1

± 0.40 °C at –15 °C
± 0.40 °C at 0 °C
± 0.50 °C at 50 °C
± 0.50 °C at 100 °C
± 0.55 °C at 120 °C

± 0.35 °C at 35 °C
± 0.50 °C at 100 °C
± 0.70 °C at 200 °C
± 1.20 °C at 350 °C
± 1.60 °C at 500 °C

Stability

± 0.10 °C at –15 °C
± 0.05 °C at 0 °C
± 0.10 °C at 120 °C

± 0.05 °C at 35 °C
± 0.20 °C at 200 °C
± 0.40 °C at 500 °C

Uniformity2 
(5.0 in dia of center 
of target)

± 0.15 °C at –15 °C
± 0.10 °C at 0 °C
± 0.25 °C at 120 °C

± 0.10 °C at 35 °C
± 0.50 °C at 200 °C
± 1.00 °C at 500 °C

Uniformity2 
(2.0 in dia of center 
of target)

± 0.10 °C at -15 °C
± 0.10 °C at 0 °C
± 0.20 °C at 120 °C

± 0.10 °C at 35 °C
± 0.25 °C at 200 °C
± 0.50 °C at 500 °C

Heating time

15 min: –15 °C to 120 °C
14 min: 23 °C to 120 °C

20 min: 35 °C to 500 °C

Cooling time

15 min: 120 °C to 23 °C 
20 min: 23 °C to –15 °C

100 min: 500 °C to 35 °C
40 min: 500 °C to 100 °C

Stabilization time

10 minutes

10 minutes

Nominal emissivity3

0.95

0.95

Thermometer emissivity compensation

0.9 to 1.0

Target diameter

152.4 mm (6 in)

Computer interface

RS-232

Power

115 V ac (± 10%), 6.3 A, 
50/60 Hz, 630 W
230 V ac (± 10%), 3.15 A, 
50/60 Hz, 630 W

115 V ac (± 10%), 10 A, 
50/60 Hz, 1000 W
230 V ac (± 10%), 5 A, 
50/60 Hz, 1000 W

Fuse(s)

115 V ac 6.3 A, 250 V, slow blow
230 V ac 3.15 A, 250 V, T

115 V ac 10 A, 250 V, fast blow
230 V ac 5 A, 250 V, F

Size (HxWxD)

356 mm x 241 mm x 216 mm 
(14 in x 9.5 in x 8.5 in)

356 mm x 241 mm x 216 mm 
(14 in x 9.5 in x 8.5 in)

Weight

9.1 kg (20 lb)

9.5 kg (21 lb)

Safety

EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04

1For 8 mm to 14 mm spectral band thermometers with emissivity set between 0.9 and 1.0. Includes stability, uniformity, emissivity, calibration and drift.
2
The uniformity specification refers to how IR thermometers with different spot sizes both focused at the center of the target will measure the same temperature.
3
The target has a nominal emissivity of 0.95, however it is radiometrically calibrated to minimize 
emissivity related uncertainties.

 

 