商品特性:
Features
Lightweight, portable, and fast
Cool to –25 °C in 15 minutes and heat to 660 °C in 15 minutes
Built-in two-channel readout for PRT, RTD, thermocouple,
4-20 mA current
True reference thermometry with accuracy to ± 0.01 °C
On-board automation and documentation
Metrology performance in accuracy, stability, uniformity, and loading
商品規格:
|
9142 |
9143 |
9144 |
Temperature Range at 23 °C |
–25 °C to 150 °C |
33 °C to 350 °C |
50 °C to 660 °C |
Display Accuracy |
± 0.2 °C Full Range |
± 0.2 °C Full Range |
± 0.35 °C at 50 °C |
Stability |
± 0.01 °C Full Range |
± 0.02 °C at 33 °C |
± 0.03 °C at 50 °C |
Axial Uniformity at 40 mm (1.6 in) |
± 0.05 °C Full Range |
± 0.04 °C at 33 °C |
± 0.05 °C at 50 °C |
Radial Uniformity |
± 0.01 °C Full Range |
± 0.01 °C at 33 °C |
± 0.02 °C at 50 °C |
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) |
± 0.006 °C Full Range |
± 0.015 °C Full Range |
± 0.015 °C at 50 °C |
Hysteresis |
0.025 °C |
0.03 °C |
0.1 °C |
Operating Conditions |
0 °C to 50 °C, 0 % to 90 % RH (non-condensing) |
||
Environmental Conditions (for all specifications except temperature range) |
13 °C to 33 °C |
||
Immersion (Well) Depth |
150 mm (5.9 in) |
||
Insert OD |
30 mm (1.18 in) |
25.3 mm (1.00 in) |
24.4 mm (0.96 in) |
Heating Time |
16 min: 23 °C to 140 °C |
5 min: 33 °C to 350 °C |
15 min: 50 °C to 660 °C |
Cooling Time |
15 min: 23 °C to –25 °C |
32 min: 350 °C to 33 °C |
35 min: 660 °C to 50 °C |
Resolution |
0.01 ° |
||
Display |
LCD, °C or °F user-selectable |
||
Size (H x W x D) |
290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in) |
||
Weight |
8.16 kg (18 lb) |
7.3 kg (16 lb) |
7.7 kg (17 lb) |
Power Requirements |
100 V to 115 V (± 10 %) 50/60 Hz, 635 W |
100 V to 115 V (± 10 %), 50/60 Hz, 1400 W |
|
Computer Interface |
RS-232 and 9930 Interface-it control software included
|